Advanced Topics¶
Creating a library IRF¶
Overview¶
Measurement of the temporal profile of an HRI for particular settings
Caution Take care to minimise the light exposure on the HRI at all times
Measurement¶
Set up for widefield imaging with HRI of interest
Use filter cube with achromatic beamsplitter with ~OD5 of absorptive ND filters
Focus light on glass slide (~4% reflection)
Record the temporal profile with 25ps gate spacings
Record 100 laser off background images with same integration time
Create library IRF file¶
Load measured IRF as FLIM data
Load camera background
Check for decay artefacts
Select small central region of image
Select Tools>Create IRF shift map…
Add to instrument calibration repository
Determining a reference lifetime¶
Load a long lifetime dye data (or a single lifetime dataset, e.g. a cell expressing GFP only)
Remove the background appropriately
IRF > Load IRF
Advanced tab > Select Use IRF shift map
Check that IRF shift map is displaying and is correct by selecting IRF shift map in the mode selector
In Fitting tab > set Fit reference to Fitted
Fit image wise both the lifetime of the sample and the reference lifetime
The reference lifetime obtained can then be used for fitting other dataset using the xml IRF shift map file as IRF